Fluke 718 -850 mbar to 6.895 bar Pressure Calibrator

ไม่พร้อมจำหน่ายในตอนนี้
ขออภัย เราไม่ทราบว่าสินค้านี้จะกลับมาในสต็อกเมื่อใด
RS Stock No.:
351-2815
หมายเลขชิ้นส่วนของผู้ผลิต / Mfr. Part No.:
FLUKE-718 100
ผู้ผลิต:
Fluke
ค้นหาผลิตภัณฑ์ที่คล้ายกันโดยเลือกคุณลักษณะอย่างน้อยหนึ่งรายการ
เลือกทั้งหมด

Brand

Fluke

Product Type

Pressure Calibrator

Battery Type

Alkaline

Maximum Pressure Measurement

6.895bar

Minimum Pressure Measurement

-850mbar

Accuracy

±0.015 %

Resolution

1 mbar

Model Number

718

Data Logging

No

Current Loop Measurement

Yes

Display Type

LCD

Power Source

Battery

Weight

992g

Minimum Operating Temperature

-10°C

Maximum Operating Temperature

55°C

Standards/Approvals

No

Width

94mm

Length

216mm

Height

66mm

Hazardous Area Certification

NO

สถานะRoHS: ได้รับการยกเว้น

COO (Country of Origin):
CN

Fluke Pressure Calibrator, ±0.015% Accuracy, 7 Bar Maximum Pressure Measurement - FLUKE-718 100


This pressure calibrator is a handheld instrument designed to provide a compact, battery-powered solution for calibrating transmitters, gauges and switches in field and workshop settings. It combines a pressure source with milliamp measurement capability and an integrated pump that resists fluid ingress, allowing technicians to generate and measure pressure while powering loop devices.

Features & Benefits


• ±0.015% mA accuracy for precise current loop verification
• Pressure range from -850 mbar to 7 bar for wide coverage
• High-resolution 0.1 mbar resolution for fine pressure discrimination
• LCD display showing simultaneous pressure and current readings
• Battery operation delivering up to 20 h typical use per set of batteries
• Integrated pump with serviceable valve assembly for field cleaning

**Applications**


• Suitable for commissioning and calibrating pressure transmitters on site
• Ideal for testing and adjusting pressure switches in process systems
• Used for functional verification of gauges across -850 mbar to 7 bar
• Can be used for loop-powered sensor checks and mA calibration
• Used with non-corrosive gases for field pressure validation

What battery type does it use and how many are required?


It uses alkaline 9V cells and requires two batteries to power all functions.

How does the pump cope with accidental fluid exposure?


The pump features protective check valves and a removable valve assembly enabling in-field cleaning using isopropyl alcohol and simple tools.

What measurement capability exists for current loops?


It measures 0 mA–24 mA with 0.001 mA resolution and an accuracy of 0.015%+1 count and can drive 20 mA into 1000 Ω when battery voltage exceeds 6.8 V.

Which pressure units and ranges are supported via the built-in pump?


The instrument supports multiple units including psi, bar, mbar, kPa, inH2O and mmHg and covers negative pressures to -850 mbar up to its full-scale positive range.

What environmental limits should users observe during operation?


Operate between -10 °C and 55 °C with specified relative-humidity limits and up to 3000m altitude

storage extends beyond these values.

Supplied with

protective yellow holster, test lead set, crocodile clips, 9V alkaline battery (fitted)

ลิงก์ที่เกี่ยวข้อง