Fluke FLK-729 300G -0.82 bar to 20.68 bar Pressure Calibrator

ไม่พร้อมจำหน่ายในตอนนี้
ขออภัย เราไม่ทราบว่าสินค้านี้จะกลับมาในสต็อกเมื่อใด
RS Stock No.:
144-5554
หมายเลขชิ้นส่วนของผู้ผลิต / Mfr. Part No.:
FLK-729 300G
ผู้ผลิต:
Fluke
ค้นหาผลิตภัณฑ์ที่คล้ายกันโดยเลือกคุณลักษณะอย่างน้อยหนึ่งรายการ
เลือกทั้งหมด

Brand

Fluke

Battery Type

Rechargeable

Product Type

Pressure Calibrator

Maximum Pressure Measurement

20.68bar

Accuracy

±0.01 %

Minimum Pressure Measurement

-0.82bar

Resolution

20.68 bar

Pressure Port Type

Differential

Model Number

FLK-729 300G

Data Logging

Yes

Current Loop Measurement

Yes

Display Type

LCD

Leak Testing

Yes

Power Source

Battery

Weight

3kg

Maximum Operating Temperature

50°C

Length

172.7mm

Standards/Approvals

No

Height

279.4mm

Hazardous Area Certification

No

สถานะRoHS: ได้รับการยกเว้น

COO (Country of Origin):
CN

Fluke Pressure Calibrator, ±0.01% Accuracy, 20.68 Bar Maximum Pressure Measurement - FLK-729 300G


This pressure calibrator is a portable, battery-powered instrument designed for pressure generation, control and measurement in process environments. It combines automatic electric pumping with pressure regulation and supports electrical loop functions including mA sourcing, simulation and measurement, enabling field technicians to perform multi-point calibrations and document results efficiently.

Features & Benefits


• Automatic pump to 20.68 bar for set-point attainment and control
• ±0.01% accuracy for precise measurement and calibration results
• 24V loop power for powering and testing transmitters
• Built-in leak testing to verify integrity of test setups
• LCD multi-channel graphical display for clear data visualisation
• Rechargeable battery operation for field portability and autonomy

Applications


• Suitable for commissioning and calibration of process transmitters
• Ideal for loop current measurement verification on plant skids
• Used for HART transmitter mA trim and light configuration tasks
• Can be used for pressure switch testing with automated sequences
• Used with RTD probes for temperature-related calibration checks

What pressure range can it cover for industrial calibrations?


It can apply and measure from -0.82 to 20.68 bar, accommodating vacuum through high-pressure tasks.

How does it simplify multi-point calibration workflows?


You enter start/end pressures and point count and the device automatically steps, stabilises and logs each test point.

What electrical measurement capabilities does it provide beyond pressure?


It measures, sources and simulates 0–24mA, measures 0–30Vdc and supplies 24V loop power for standalone tests.

How does the instrument handle minor system leaks during a test?


An automatic fine-pressure adjustment feature compensates for small leaks to maintain set-points without manual vernier adjustments.

What environmental operating limits affect pressure control and charging?


Measurement functions operate from -10°C to 50°C while pressure control is rated 0°C to 50°C and the battery will only charge up to 40°C.

ลิงก์ที่เกี่ยวข้อง