IP66 rated Measuring range up to 10m Unaffected by temperature, vacuum, pressure, vapours, gases or density Non-contact measurement with microwave pulses Accurate to 5mm using Echofox technology Maintenance-free rugged design Display/adjustment in head via PLICSCOM module or VEGADIS 61 remote housing (both types not supplied with unit) 2-wire (20 to 36V loop powered) technology; HART protocol with universal commands Output signal: 0//4 to 20mA (HART) 3/4in. BSP mount Vessel pressure:; -1 to +3 bar Housing material: PBT Transducer material: PVDF Cable entries: 2 x M20 Simple calibration and in-head display of echo curve Fully communicative built-in data and trend logging
Note: PLICSCOM module and VEGADIS 61 remote housing available separately – see stock nos. 468-3514 and 468-3508 respectively
Note
This product must be licensed by the user and is supplied with a licensing application form. Licensing is free of charge. Users should contact RS Technical in case of difficulty and for clarification of requirements.